6850 Series GC Site Preparation
Table 37
Gases usable with Agilent GCs and packed columns
Detector type
Electron capture (ECD)
Flame ionization (FID)
Flame photometric (FPD)
Thermal conductivity (TCD)
1 Slightly greater sensitivity than helium. Incompatible with some compounds.
2 For analysis of hydrogen or helium. Greatly reduces sensitivity for other compounds.
For installation checkout, Agilent requires the gas types shown in
82
Carrier gas
Comments
Nitrogen
Maximum sensitivity
Argon/methane
Maximum dynamic range
Nitrogen
Maximum sensitivity
Helium
Acceptable alternative
Hydrogen
Helium
Nitrogen
Argon
Helium
General use
Hydrogen
Maximum sensitivity
Nitrogen
Hydrogen detection
Argon
Maximum hydrogen
sensitivity
Detector, anode purge, or
reference
Nitrogen
Argon/Methane
Hydrogen and air for
detector.
Hydrogen and air for
detector.
Reference must be same
as carrier and makeup.
1
2
1
GC, GC/MS, and ALS Site Preparation Guide
Table
38.