6850 Series GC Site Preparation
Table 28
Table 28
Tool
Custom Tee, G3430-60009
ECD/TCD Detector plug, 5060-9055
1/8-inch Ball Valve, 0100-2144
Digital flow meter, Flow tracker 1000
Electronic gas leak detector (G3388B)
Column cutters
T-10 and T-20 Torx drivers
1/8-inch tubing cutter (wire cutter type)
Assorted wrenches: 1/4-inch, 3/8-inch,
7/16-inch, 9/16-inch
Electronic vial crimper
Table 29
should consider purchasing the following supplies in order to maintain their
system and prevent interruptions in the use of their system. Please refer to the
latest Agilent consumables and supplies catalog and to the Agilent web site at
www.agilent.com/chem
periods.
64
lists other useful tools not included with the GC.
Useful tools not included with GC
lists consumables that you may wish to order. First time GC users
for part numbers and recommended maintenance
Used for
Connecting the same gas to front and back EPC
module.
Inlet pressure decay test
Inlet pressure decay test (one per inlet)
Verifying flows, checking for leaks and plugs
Locating gas leaks; safety checks when using
Hydrogen
Cutting columns
Removing tray; removing covers to access EPC
modules, traps, and possible leaks
Cutting gas supply tubing
Gas supply and plumbing fittings
Assuring consistently air-tight vial closure,
regardless of who does the crimping
GC, GC/MS, and ALS Site Preparation Guide