7890 Series GC Site Preparation

Environmental Conditions

Operating the instrument within the recommended ranges optimizes
instrument performance and lifetime. Performance can be affected by sources
of heat and cold from heating, air conditioning systems, or drafts. See
The conditions assume a noncondensing, noncorrosive atmosphere. The
instrument meets the following International Electrotechnical Commission
(IEC) classifications: Equipment Class I, Laboratory Equipment, Installation
Category II, and Pollution Degree 2.
Table 6
Environmental conditions for operation and storage
Product
7890 Series GC
MSD
5975 Series MSD
5977 Series MSD
MS
7010 or 7000 Triple Quad
MS
7200 or 7250 Q-TOF MS
28
Conditions
Standard oven ramp
Fast oven ramp (options 002 and 003)
Storage
Operation
Storage
Operation
Storage
Operation
Storage
Operation
Storage
Operating temp
Operating
range
humidity range
15 to 35 °C
5 to 95%
15 to 35 °C
5 to 95%
–40 to 70 °C
5 to 95%
1
15 to 35 °C
20 to 80%
(59 to 95 °F)
–20 to 70 °C
0 to 95%
(–4 to 158 °F)
1
15 to 35 °C
20 to 80%
(59 to 95 °F)
–20 to 70 °C
0 to 95%
(–4 to 158 °F)
2
15 to 35 °C
40 to 80%
(59 to 95 °F)
–20 to 70 °C
0 to 95%
(–4 to 158 °F)
2
15 to 35 °C
20 to 80%
(59 to 95 °F)
–20 to 70 °C
0 to 95%
(–4 to 158 °F)
GC, GC/MS, and ALS Site Preparation Guide
Table
6.
Maximum
altitude
4,615 m
4,615 m
4,615 m
4,615 m
3
5,000 m
2,500 m