6850 Series GC Site Preparation
Table 32
Power requirements (continued)
Product
Oven type
5975 Series MSD
5975 Series MSD
5977 Series MSD
5977 Series MSD
5977 Series MSD
All
Data system PC
(monitor, CPU, printer)
Do not use extension cords with Agilent instruments. Extension cords normally are
WA R N I N G
not rated to carry enough power and can be a safety hazard.
Although your GC should arrive ready for operation in your country, compare
its voltage requirements with those listed in
ordered is not suitable for your installation, contact Agilent Technologies.
Note that ALS instruments receive their power from the GC.
A proper earth ground is required for GC operations. Any interruption of the grounding
C A U T I O N
conductor or disconnection of the power cord could cause a shock that could result in
personal injury.
74
Line voltage (VAC)
Frequency
(Hz)
220–240
50/60 ± 5%
(–10% / +5%)
200
50/60 ± 5%
(–10% / +5%)
120
50/60 ± 5%
(–10% / +5%)
220–240
50/60 ± 5%
(–10% / +5%)
200
50/60 ± 5%
(–10% / +5%)
100/120/200-240
50/60 ± 5%
(–10% / +5%)
Maximum
Current
continuous power
rating
consumption (VA)
(amps)
1100 (400 for
8
foreline pump only)
1100 (400 for
8
foreline pump only)
1100 (400 for
8
foreline pump only)
1100 (400 for
8
foreline pump only)
1100 (400 for
8
foreline pump only)
1000
15
Table
32. If the voltage option you
GC, GC/MS, and ALS Site Preparation Guide
Power outlet
current rating
10 Amp
Dedicated
10 Amp
Dedicated
10 Amp
Dedicated
10 Amp
Dedicated
10 Amp
Dedicated
15 Amp
Dedicated