6850 Series GC Site Preparation
Table 42
Hydrogen
Air
Makeup
Reference
Requirements for hydrogen as a carrier gas
Hydrogen can be supplied from a generator or from a cylinder.
Agilent recommends use of a high-quality hydrogen gas generator. A
high-quality generator can consistently produce purity > 99.9999%, and the
generator can include built-in safety features such as limited storage, limited
flow rates, and auto-shutdown. Select a hydrogen generator that provides low
(good) specifications for water and oxygen content.
If using a hydrogen gas cylinder, Agilent recommends use of Gas Clean Filters
to purify the gas. Consider additional safety equipment as recommended by
your company safety personnel.
86
Delivery pressures for detectors required at the GC/MS, in kPa (psig)
Detector type
FID
240–690 (35–100)
380–690 (55–100)
380–690 (55–100)
Conversions: 1 psi = 6.8947 kPa = 0.068947 Bar = 0.068 ATM
TCD
ECD
380–690 (55–100)
380–690 (55–100)
380–690 (55–100)
GC, GC/MS, and ALS Site Preparation Guide
FPD
310–690 (45–100)
690–827
(100–120)
380–690 (55–100)