Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

7820 MSD Site Preparation

Environmental Conditions

Operating the instrument within the recommended ranges optimizes
instrument performance and lifetime. Performance can be affected by sources
of heat and cold from heating, air conditioning systems, or drafts. See
Table
The instrument meets the following International Electrotechnical
Commission (IEC) classifications: Equipment Class I, Laboratory Equipment,
Installation Category II, and Pollution Degree 2.
Table 56
Environmental conditions for operation and storage
Product
7820A GC
MSD
5975 Series MSD
5977 Series MSD
* Operation requires constant temperature (variations < 2
118
56. The conditions assume a noncondensing, noncorrosive atmosphere.
Conditions
Standard oven ramp
Storage
Operation
Storage
Operation
Storage
Operating temp
range
5 to 45 °C
–20 to 65 °C
*
15 to 35 °C
(59 to 95 °F)
–20 to 70 °C
(–4 to 158 °F)
*
15 to 35 °C
(59 to 95 °F)
–20 to 70 °C
(–4 to 158 °F)
o
C/hour)
GC, GC/MS, and ALS Site Preparation Guide
Operating
Maximum
humidity range
altitude
5 to 90%
3,100 m
0 to 90%
20 to 80%
4,615 m
0 to 95%
20 to 80%
4,615 m
0 to 95%

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

MsGc

Inhaltsverzeichnis