6850 Series GC Site Preparation

Environmental Conditions

Operating the instrument within the recommended ranges optimizes
instrument performance and lifetime. Performance can be affected by sources
of heat and cold from heating, air conditioning systems, or drafts. See
Table
The instrument meets the following International Electrotechnical
Commission (IEC) classifications: Equipment Class I, Laboratory Equipment,
Installation Category II, and Pollution Degree 2.
Table 35
Environmental conditions for operation and storage
Product
6850 Series GC
MSD
5975 Series MSD
5977 Series MSD
1 Operation requires constant temperature (variations < 2
80
35. The conditions assume a noncondensing, noncorrosive atmosphere.
Conditions
Standard oven ramp
Fast oven ramp (options 002 and 003)
Storage
Operation
Storage
Operation
Storage
Operating temp
range
15 to 35 °C
15 to 35 °C
–5 to 40 °C
1
15 to 35 °C
(59 to 95 °F)
–20 to 70 °C
(–4 to 158 °F)
1
15 to 35 °C
(59 to 95 °F)
–20 to 70 °C
(–4 to 158 °F)
o
C/hour)
GC, GC/MS, and ALS Site Preparation Guide
Operating
Maximum
humidity range
altitude
5 to 95%
4,615 m
5 to 95%
4,615 m
5 to 95%
20 to 80%
4,615 m
0 to 95%
20 to 80%
4,615 m
0 to 95%