K5+ – Troubleshooting
The computer interrupts the job and displays
that the vacuum is too low
You can resolve this problem while the job
execution is interrupted. DENTALCNC
continues the job as soon as the problem is
resolved.
Is the suction unit switched on and operating?
How to check this:
1. Check if the suction hose is properly seated in the
corresponding opening and is undamaged.
2. If the machine controls the suction unit:
a. Check if the switching unit or data cable is
properly installed.
b. Try operating the suction unit without the
switching unit or data cable.
3. Switch on the suction unit.
4. Set the extraction level higher until the machine
continues the job.
Is the filter or container of the suction unit full?
If applicable:
Insert an empty filter into the suction unit or empty
the container.
Does the suction unit have an automatic
tapping function?
If applicable:
Choose a shorter tapping interval.
I have exchanged the tool magazine inserts
and now there are no holes for the tools
anymore
Tool magazine inserts are delivered without drilled
holes for tools. You drill them with the machine.
Use DENTALCNC to drill new holes into the new
inserts.