Gently wipe surface of probe lens window with cotton swab or
cloth slightly moistened with isopropyl or ethyl alcohol only.
Note: Use gentle pressure when cleaning the sensor to
prevent damaging the unit by accidently changing the
position of the sensor.
With probe facing down, wipe the probe with a damp cloth or
cleaning wipe moistened with isopropyl or ethyl alcohol.
Gently wipe with dry, clean cotton swab or cloth immediately.
Allow at least 5 minutes drying time before taking temperature.
Make sure probe lens window is clean and dry prior to use.

Cleaning the contacts

Caution: Do not use bleach solutions of any kind when cleaning metal electrical contacts.
It will damage the device.
Slightly dampen a cotton swab with 70% isopropyl alcohol.