Remove probe cover box carrier from the cradle by rotating the
probe cover box carrier forward.
Place probe cover box carrier back in cradle by lining up the
brackets and pushing down.
14.6 Storage instructions
Store thermometer and probe covers in a dry location (the thermometer is not protected against
ingress of water), free from dust and contamination and away from direct sunlight. The ambient
temperature at the storage location should remain fairly constant and within the range of 50 °F to
104 °F (10 °C to 40 °C) to allow for immediate use.
14.7 Replacing the batteries
The thermometer is supplied with two 1.5 V batteries type AA (LR 6).
For best performance, Duracell
Note: The battery life performance testing was based on the Duracell
other than these are not guaranteed to provide the same life performance results.
Insert new batteries when the battery
symbol begins to flash on the display
(See 12. Errors and notifications).
Open battery door by sliding spring
latch to the right using your finger
or a pointed object such as a pen.
While holding the latch in the open
position, grasp the battery door and
remove it.
alkaline batteries are recommended.
®
alkaline batteries. Batteries
®
27