14. Maintenance and service

14.1 Cleaning the probe lens window, probe and contacts
Only use Braun ThermoScan
Do not use damaged, perforated, soiled or poor fi tting probe covers. Do not reuse probe covers.
Dirty probe lens window = lower readings. Fingerprints, cerumen, dust and other soiling
components reduce transparency of the tip and result in lower temperature measurements.
If thermometer is placed in ear without a probe cover, clean immediately.
Do not damage probe lens window. Avoid touching probe lens window except when cleaning.
If probe lens window is damaged return to Welch Allyn for service.
Do not modify, change or adjust probe lens window. These changes will affect calibration and
accuracy of the thermometer. If probe lens window is damaged return to Welch Allyn for service.
Do not use cleaning solution other than isopropyl or ethyl alcohol to clean probe lens
window and probe. Bleach and other cleaning agents will cause permanent damage to the
probe and probe lens window.
Cleaning the probe lens window and probe
The thermometer probe and probe lens window should be cleaned when there are fi ngerprints,
cerumen, dust or other soiling components present, following the directions below:
Remove probe cover and discard.
Moisten cotton swab or cloth slightly with isopropyl or ethyl
alcohol. Do not saturate.
disposable probe covers.
®
21