Herunterladen Inhalt Inhalt Diese Seite drucken

Error Message - Braun Welch Allyn Thermoscan PRO 6000 Gebrauchsanweisung

Ohrthermometer
Vorschau ausblenden Andere Handbücher für Welch Allyn Thermoscan PRO 6000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 634
12. Errors and notifications

Error message

Situation
No probe cover is attached
(animates ON).
Used probe cover is attached
(animates OFF).
(POS = position error)
The infrared monitor cannot find
a temperature equilibrium and
allows no measurement.
Ambient temperature is not
within the allowed operating
range (50–104 °F or 10–40 °C)
or changing too rapidly.
Temperature taken is not within
typical human temperature
range.
HI will be displayed when
temperature is higher than
108 °F (42.2 °C).
LO will be displayed when
temperature is lower than 68 °F
(20 °C).
Solution
Attach new, clean probe cover.
Discard probe cover that is on
and attach new, clean probe cover
if taking another temperature
measurement.
Change the probe cover to reset.
Restrict patient movement and
ensure that the positioning of the
probe is correct and remains stable
while taking new temperature.
Wait 20 sec. until thermometer
turns off automatically, then turn
on again. Ensure thermometer and
patient are in an environment for
30 minutes where the temperature
is between 50 ºF and 104 ºF
or 10 ºC and 40 ºC prior to
measurement.
Change probe cover to reset.
Then, make sure thermometer is
properly inserted and take a new
temperature.