Slide switch to C or F using a pen or pointed object.
Place batteries back into thermometer. Snap the battery door
back into place and ensure the latch returns to the original
latched position. Celsius or Fahrenheit symbol will appear on the
display.

11.2 Advanced functions

The Welch Allyn Service tool software is required to modify the thermometers configuration.
A charging station and rechargeable batteries or compatible Welch Allyn Device is required to
connect to the PC running the Welch Allyn Service Tool. (See 11.3 Advanced function settings
and 11.4 Service tools)
Item
PerfecTemp™
C/F button
Default C/F manual
switch
Security function
Timer icon
Unadjusted
operating mode
Description
Improves reading accuracy by
detecting placement of probe in
ear canal
Use the C/F button to view
measurements in the non‑default
(alternate) temperature scale.
When off (disabled) only the
default scale is available.
When on (enabled), the default
scale can be set using the
manual switch in the battery
compartment. When the control
is off (disabled), the radio buttons
for Celsius and Fahrenheit are
enabled allowing the service tool
to set the default scale.
Sets countdown time after
removal from charging dock to
lock
Displays an icon along with the
timer counter
Places thermometer in mode to
detect raw ear temperature only
Settings
On/Off
On/Off
On/Off
1 through 12 hours
On/Off
Allows user to set
device to unadjusted
operating mode
Default
setting
On
On
On
Off
On
Off
15