Settings
56
After the Save button is pressed, the gradient factor for
clockwise and anticlockwise torque is calculated and dis
played in the table. Then activate Supporting points
approximation for low-pass filter 1 and/or 2 with ON, or
deactivate this with OFF.
Clockwise torque = positive torque
Anticlockwise torque = negative torque
Example 2
Input interpolation equation
Select the channel to which Supporting points approxima
tion should be applied, and now select Interpolation
gradient from the Interpolation points equation menu
item. Pressing the Save button adopts the entered gradi
ent factor. Then activate Supporting points approximation
for low-pass filter 1 and/or 2 with ON, or deactivate it with
OFF.
If you have a calibration certificate for the connected
torque flange, you can take these parameters directly
from the calibration certificate.
Gradient factors
A3704-4.0 HBM: public
TIM-EC