Error specification
Error display time
Manual error flag time (in ms)
Valid range: 0 to 20000 ms
TIM-EC
®
CAT
master. See also Chapter 11.3
Parameterizing process data
You can specify how the measurement system should
react to incorrect measured values under Error settings.
The following settings are possible:
Error settings
If, for instance, the torque value is incorrect (e.g. actual
value Wert > 120 % of the rated value or transmission
error between rotor and stator), then the output will con
form to what you have set under error specification. If you
selected Overflow, the torque value will change to
7FFFFFFFh in the case of an error.
The time how long the error should be signalled on the
®
EtherCAT
interface can be set manually or automati
cally determined by the system.
If you select automatic, the error duration for dynamic
errors will be set dependent on the set filter. Static errors
are displayed until they are rectified.
A3704-4.0 HBM: public
Overflow (7FFFFFFFh)
Underflow (80000000h)
Last valid value
Actual value
automatic; manual
0
Settings