Herunterladen Inhalt Inhalt Diese Seite drucken

Battery Charger - Otto Bock C2000 Gebrauchsanweisung

Frontantrieb/front-wheel drive;heckantrieb/rear-wheel drive
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation

6.6.2 Battery Charger

WARNING
Explosion hazard due to sparks. Always switch off the
battery charger and disconnect it from the power source
before disconnecting the battery.
NOTICE
Unauthorised battery replacement. The battery may only
be replaced by authorised specialist staff. The characteris-
tic curve of the battery charger set at the factory corre-
sponds to the battery provided and must not be changed.
Setting the characteristic curve incorrectly can result in
permanent damage to the battery.
NOTICE
Risk of damage to the battery charger / risk of damage
caused by the battery charger. Please note the following
when using the charger:
n Only use a battery charger supplied by Ottobock which
has been tested and approved by Ottobock for the re-
spective batteries (observe information on the charger).
Failure to do so can result in a battery explosion and
possible impairment of health due to contact with bat-
tery acid.
156
Ottobock
|
n The information on the nameplate of the charger must
match the country-specific voltage of the respective
power supply network.
n Only use the battery charger within the specified rang-
es of temperature and humidity.
n Place the rubber feet of the battery charger on a level
surface.
n When setting the charger up close to a window, protect
it against direct sunlight.
n Keep the charger from overheating. The vent openings
on the back of the charger must not be covered.
n Turn off the control unit during the charging process so
that the entire charging current flows to the battery.
n Avoid dust and dirt. Only clean with a dry cloth.
The battery charger is designed for the 12 V gel batteries
installed in the C2000 power wheelchair. The charger can
operate along two characteristic curves to charge the re-
spective batteries. Upon delivery of the power wheelchair,
the matching characteristic curve will be set at the factory. If
the battery charger is to be used with another power wheel-
chair or if batteries with another capacity are installed, the
characteristic curve has to be checked.