C C A A U U T T I I O O N N - - Always raise the IV pole before you attach the defibrillator tray/foot extender to the product. If you do not raise
the IV pole, the foot extender will not operate.
To convert the defibrillator tray/foot extender to a foot extender (Figure 11):
1. Pull out the top knob (A).
2. Pivot the defibrillator tray (B) until the tray locks against the foot extender.
3. Pull out the bottom knob (D) while you hold the defibrillator tray/foot extender assembly.
4. Lower the foot extender (C) until the foot extender is flat.
5. Release the bottom knob (D). Push on the foot extender to make sure that the foot extender is locked in place.
N N o o t t e e
• Do not use the defibrillator tray/foot extender as a push/pull device. Product damage may occur.
• Do not attach items to the foot extender.
A A t t t t a a c c h h i i n n g g t t h h e e f f o o o o t t b b o o a a r r d d / / c c h h a a r r t t h h o o l l d d e e r r
W W A A R R N N I I N N G G - - Always use caution if the defibrillator tray/foot extender, footboard/chart holder, or upright oxygen bottle
holder is attached to avoid pinching your fingers when you position the foot end push handle option.
To attach the footboard/chart holder, insert the footboard/chart holder pins into the sockets at the foot end of the product.
N N o o t t e e - - Do not use the footboard/chart holder as a push/pull device. Product damage may occur.
P P o o s s i i t t i i o o n n i i n n g g t t h h e e t t w w o o - - s s t t a a g g e e p p e e r r m m a a n n e e n n t t l l y y a a t t t t a a c c h h e e d d I I V V p p o o l l e e
W W A A R R N N I I N N G G - - Do not use the IV pole as a push/pull device. Product damage may occur.
You can purchase the product with the two-stage IV pole option permanently attached at the head end, foot end, or both
ends of the product. The IV pole is equipped with a telescopic pole that extends to provide a second height position. You
can fold and store the IV pole when not in use.
To position the two-stage IV pole (Figure 12):
1. Lift and pivot the pole from the storage position.
2. Push the IV pole down until the IV pole locks in place.
3. To raise the height of the IV pole, pull up on the telescoping portion (A) until the pole locks in place at the fully raised
position.
4. Rotate the IV hangers (B) to the desired position and hang the IV bags.
5. To lower the IV pole, hold the telescoping portion of the IV pole, turn the latch (C), and lower the telescoping portion.
N N o o t t e e
• Do not hang IV bags that exceed 40 lb (18 kg) on the IV pole.
• Always make sure that the IV pole is at a low height to pass safely through door openings when you transport a
patient.