To lower the Fowler backrest, squeeze the Fowler release handles (A) and push the Fowler backrest down to the desired
position (Figure 3 ).
R R a a i i s s i i n n g g o o r r l l o o w w e e r r i i n n g g t t h h e e F F o o w w l l e e r r b b a a c c k k r r e e s s t t o o r r t t h h e e G G a a t t c c h h w w i i t t h h t t h h e e c c r r a a n n k k o o p p t t i i o o n n
To raise the Fowler backrest, turn the crank handle clockwise.
To lower the Fowler backrest, turn the crank handle counterclockwise.
To raise the Gatch, turn the crank handle clockwise.
To lower the Gatch, turn the crank handle counterclockwise.
N N o o t t e e - - The Fowler backrest and Gatch crank handles are stored under the litter. Pivot the crank out and push in to secure
the crank rod.
P P o o s s i i t t i i o o n n i i n n g g t t h h e e e e n n h h a a n n c c e e d d c c l l e e a a r r a a n n c c e e h h e e a a d d p p i i e e c c e e
W W A A R R N N I I N N G G
• Always support the patient's head when you position the headpiece or the Fowler backrest. Patient injury may occur.
• Do not reach between the side of the head extension and the articulating headpiece to pull the release handles.
Operator injury may occur.
• Always keep fingers away from jointed areas when you adjust the headpiece. Operator injury may occur.
To adjust the articulating headpiece, grasp either handle under the headpiece and squeeze.
Handle (A) (Figure 5) releases one latch and rotates the headpiece on axis (C) (Figure 4).
Handle (B) (Figure 5) releases the other latch and rotates the headpiece on axis (D) (Figure 4).
EN
12
A
F F i i g g u u r r e e 3 3 – – F F o o w w l l e e r r b b a a c c k k r r e e s s t t
A
1089-109-005 Rev AA.0