Error
Error Name
03
Liquid
04
File missing
05-09
N/A
10
Over temperature Board
11
Over temperature Board
12
Over temperature Tank
13
Over temperature Tank
14
Tank temperature rising
15
Tank temperature rising
20
Overcurrent
21
Overcurrent
22
Overvoltage
23
Overvoltage
24
Power deviation
25
Power deviation
26
Power deviation
Description
Liquid level is too low
Firmware files are missing
N/A
Temperature on power PCB
assembly above 60ºC/140ºF
Hardware has shutdown due
to temperature on power
PCB assembly above
65ºC/150ºF
Tank solution temperature is
too high
Tank solution temperature is
too high
Solution bath temperature is
increasing too quickly
Solution bath temperature is
increasing too slowly
Abnormal current level
Abnormal current level
Abnormal voltage level
Abnormal voltage level
Power output is lower than
expected
Power output is lower than
expected
Power output is higher than
expected
Recommended User Action
Verify that the solution level in the tank is filled
between the minimum and maximum fill lined
stamped on the inside back of the tank. If
solution level is withing the fill lines, contact
customer service.
Contact customer service for firmware update.
N/A
Ensure room temperature and humidity are
within the recommended operational
environmental conditions and there is at least
2 inches of clearance on all sides of the unit for
adequate ventilation. If problem persists,
contact customer service.
Ensure room temperature and humidity are
within the recommended operational
environmental conditions and there is at least
2 inches of clearance on all sides of the unit for
adequate ventilation. If problem persists,
contact customer service.
If the unit has been operated for extended
periods of time (e.g. multiple hours without
intermission) the bath temperature may rise
beyond an acceptable level. Either change the
solution or wait for it to cool. If the problem
persists, contact customer service.
If the unit has been operated for extended
periods of time (e.g. multiple hours without
intermission) the bath temperature may rise
beyond an acceptable level. Either change the
solution or wait for it to cool. If the problem
persists, contact customer service.
Ensure that the tank is filled with solution to
within the fill lines. If problem persists, contact
customer service.
Ensure that the tank is filled with solution to
within the fill lines. If problem persists, contact
customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.
Verify that the input voltage to the unit is
within the specifications noted in the
instructions for use. Deviations in supply
voltage may trigger this error. If the problem
persists, contact customer service.