• Always rinse well with clear water (2 or 3 times) after cleaning. Wipe dry completely.
• Specialized household stainless steel cleaner may also be used.
 Avoid:
• Do not let dirt accumulate.
• Steel or stainless steel wool or scrapers to remove stubborn debris.
• Harsh or abrasive cleaners.
• Chemicals harmful to stainless steel not limited to but including products that contain
chloride, fluoride, iodide or bromide.
• Cleaners that contain bleach.
In case of questions regarding compatibility of decontamination or cleaning agents please contact
Coltène/Whaledent, Inc.
Inspecting the Power Cord Periodically
Check the power cord for damage and confirm that the connectors at both ends of the cord are
fully seated.
Changing the Fuse
• Remove the power cord from the wall outlet first and then from the power cord receptacle on
the rear of the unit.
• The fuse compartment is located directly above to the power cord receptacle. To access, take a
small, flat-headed screwdriver to the tabs on the fuse holder. Apply slight pressure to release
(Figure
the fuse holder
• Gently remove the fuses from the fuse holder.
(Figure
• Examine the fuses
current. If the fuse appears to be in good condition, return the unit to your dealer or Coltène/
Whaledent. This is an indication that the unit is defective.
• Plug the power cord into the power cord receptacle on the rear of the unit. Then plug the power
cord into the wall receptacle.
NOTE: The fuse is a protection device designed to limit the amount of current being drawn or
supplied to the unit. If a fuse is blown, this can mean that there was a surge of current to the unit
due to an external power surge such as lightning, or more likely due to an internal component
failure of the cleaner.
*see last page for figures
9*).
10*). If blown, replace with a fuse that matches the unit's voltage and