Firmware Update Process
Checking the Firmware Version
• Press the "Setup" icon from the main menu, bottom left icon on the display.
• Press the "Information" icon from the setup menu, far right icon on the display.
• The unit serial number and firmware version will be displayed.
• Press the "Check Mark" icon then the "Home" icon to return to the main menu.
Firmware Update Process
• Place the most current firmware update folder "coltene-update" on a blank USB stick.
• With the BioSonic UC150 displaying the main menu, place the USB stick in the display's USB port.
• Wait a minute and the software update screen will appear on the display noting the current.
firmware version on the BioSonic UC150 and the current version on the USB stick.
• Press the "Update" icon to start the software update process.
• Do not touch the unit or the USB stick during the update process until the unit reboots and the
main menu is displayed.
• The firmware update process is complete and the USB stick can be removed.
13.
Troubleshooting
NOTE: Consult your dealer or Coltène/Whaledent for service if needed.
Error Codes
• The BioSonic UC150 is designed to interrupt a cleaning cycle and display an error code if a
problem is detected. The table below lists the BioSonic UC150 error messages and the
recommended user actions if encountered.
NOTE: The below table is provided as a reference to help users further understanding the error
codes of the BioSonic UC150.
Error
Error Name
00
No Error
01
Aborted by user
02
Power interrupted
16
Customer Error Code Table
Description
Cycle completed
Cycle was aborted by the
user
Power to (or within) the unit
was interrupted
Recommended User Action
Displayed in the user log file
This error is displayed in the user log file only
to note that the cleaning cycle was not
completed. The unit can continue to be used
normally.
If power was interrupted by external means
(power outage, unit was unplugged, display
was removed) press the "OK" icon on the
display screen to go to the main menu and
resume the next cleaning cycle.