Bedienungsanleitung Software
PLC --> ME
Address
+92.0
+96.0
+100.0
+104.0
+108.0
+112.0
+116.0
+120.0
+124.0
PLC <-- ME
Address
+0.0
+0.1
+0.2
+0.3
+0.4
+0.5
+0.6
+0.7
+1.0
+1.1
+1.2
+1.3
+1.4
+1.5
+1.6
+1.7
+4.0
+8.0
+12.0
+16.0
+20.0
thicknessCONTROL RTP 8301.EO / 4350039.85
Name
Name
Alive Signal
Centralized Fault
System Ready
Measurement Active
Out of Tolerance
Sensor Dirty
Axis in Parking Position
Calibration Active
Fix Position Measurement
Active
16 Bit Reserve
Current Thickness
Current Position
Current Length
Used Setpoint Thickness
Thickness Track 1
Type
Type
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
Bit
REAL
REAL
REAL
REAL
REAL
Description
Description
Alive Signal: Pulse with
about 1 Hz.
0: no error
1: system fault, e.g. motor
fault, ... (details will be
shown in GUI)
1: system ready for measu-
rement
1: measurement is active
1: meanvalue of cross pro-
file is out of limits
1: sensor is dirty, please
clean - can only be che-
cked during calibration
1: axis in parking position
1: calibration in process
1: measurement in fix posi-
tion is active
current thickness [mm]
current position of sensor
head [mm]
current length [m] (calcula-
ted by rotation of roller)
used setpoint for thickness
[mm]
thickness of track 1 [mm]
(position defined in system
parameters)
Seite 55