Herunterladen Inhalt Inhalt Diese Seite drucken

Product Description; Fields Of Application; Technical Data; Charger - 3M Elipar DeepCure-S Betriebsanleitung

Led-lichtgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
SEITE 3 - 148 x 210 mm - 3100021789/01 - SCHWARZ - 20-053 (kn)
Light guide
Glare Shield

Handpiece

Product Description

Elipar DeepCure-S is a high-performance LED light
source for polymerization of dental materials. It consists
of a charger and a wireless handpiece powered by a
rechargeable battery. The device is a medical electrical
device in accordance with IEC 60601-1 and is available
as a tabletop device. Wall mounting is not possible.
In comparison with conventional light-curing devices,
Elipar DeepCure-S features excellent beam collimation
and a uniform beam profile, directing more of the light
energy to the restoration being polymerized and
producing a deep, uniform, and complete curing of the
restoration.
The light source is a high-performance light diode (LED).
The beam emerging from the device covers the light
wavelength range of 430 to 480 nm relevant, for
instance, for camphor quinone products and is suitable
for use with the majority of light-curing dental materials,
including materials for fillings, liners, core build-ups,
fissure sealings, temporary restorations, and cements
for indirect restorations.
See the manufacturer's information for the exposure
time required for the specific dental material.
Settable exposure times:
• 5, 10, 15, 20 sec
• Continuous mode (120 sec)
• Tack-cure mode (1 sec)
The charger is equipped with an integrated light intensity
meter for the Elipar DeepCure-S.
The unit is shipped with a light guide with 10 mm
diameter. It is not permissible to use the light guides of
other units.
The handpiece is equipped with a sleep mode to mini-
mize the unit's energy consumption. The handpiece
switches to sleep mode once it is placed in the charger or
Power cord
if left unused for approx. 5 minutes outside the charger.
+
These Instructions for Use should be kept for the
duration of product use. The product may only be
used when the product labeling is clearly readable.
For details on all additionally mentioned products
please refer to the corresponding Instructions for
Use.
Intended Purpose
Intended users: educated dental professionals, e.g.

Charger

general dentists, dental assistants, dental hygienists,
who have theoretical and practical knowledge on usage
of dental products.

Fields of Application

• Polymerization of light-curing dental materials with
photo initiator for the wavelength range 430-480 nm.
- Though the majority of light-curing dental materials
are responsive in this range of wavelengths,
you may wish to contact the manufacturer of the
material in question.

Technical Data

Charger
Operating voltage:
Power input:
Dimensions:
Weight:
Classification:
Handpiece
Power supply:
Utilizable wavelength
range:
Wavelength peak:
100-127 V 50/60 Hz
230 V 50/60 Hz (see rating
plate for factory-set voltage)
0.08 A (230 V)
0.16 A (100-127 V)
Length: 170 mm
Width:
95 mm
Height:
50 mm
650 g
Protection class II,
lithium-ion battery, nominal
voltage 3.7 V
430-480 nm
444-452 nm
3

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Diese Anleitung auch für:

122629 configurable

Inhaltsverzeichnis