Clean Charger, Handpiece, And Glare Shield - 3M Elipar DeepCure-S Betriebsanleitung

Led-lichtgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
SEITE 11 - 148 x 210 mm - 3100021789/01 - SCHWARZ - 20-053 (kn)
®
Enzol
and CaviWipes™ were used for manual clean-
ing.
2
Cidex OPA, ready-for-use solution, was used for vali-
dation of the high level disinfection and CaviWipes™
for validation of intermediate level disinfection.
Packaging for Sterilization
Package the light guide in single wrapped sterilization
pouches (one-way packaging or double packaging).
Sterilization
Effective cleaning and disinfection are absolutely essen-
tial requirements for effective sterilization.
A fractionated vacuum procedure (at least 3 vacuum
stages) or a gravity procedure (with adequate product
drying) must be used for the sterilization. The use of the
less effective gravity procedure is permissible solely if
the fractionated vacuum procedure is unavailable. The
user is solely responsible for selection of the parame-
ters required for the gravity procedure; the user must
therefore calculate or validate these parameters in ac-
cordance with the loading configuration. The drying time
actually required is also directly dependent on parame-
ters that are in the sole responsibility of the user (load-
ing configuration and density, sterilizer condition, etc.)
and must consequently be determined by the user.
• Steam sterilizer in conformity with DIN EN 13060/DIN
EN 285 or ANSI AAMI ST79 (for USA: FDA clearance)
• Maximum sterilization temperature 135 °C (275 °F)
• The following cycles were validated for the light
guide:
Fractional prevacuum
132 °C
Temperature
(270 °F)
Sterilization
4 min
time
Drying time 30 min
Check
Before using the light guide again, check it for damaged
surfaces, discoloration, and contamination; do not use
damaged light guides. If the light guide is still contami-
nated, repeat all reprocessing steps.
Clean and Disinfect Handpiece and Glare Shield
• Cleaning and disinfecting solutions must not enter the
unit.
Gravity dis-
placement
134 °C
121 °C
(273 °F)
(250 °F)
3 min
20 min
30 min
30 min
• Wipe the surfaces with ready-to-use disinfectant
wipes3 (e.g., CaviWipes™ based on the active sub-
stances alcohols and quaternary compounds (QAV) or
equivalent disinfection agents). Subsequently inspect
the surfaces for visible impurities. If any impurities
are visible, repeat the cleaning procedure.
• Solvents or abrasive cleaners may not be used as
they can damage the components.
• Wipe the device surfaces with a tuberculocidal disin-
fectant wipe3 for disinfection and keep it wet accord-
ing to the disinfection agent manufacturer's direc-
tions for use, particularly observing the contact time.
• Disinfectants must not be sprayed directly onto the
device surface.
• Wipe the device surfaces with a soft cloth moistened
with water (drinking water quality as a minimum) and
dry with a soft, lint-free cloth. If you do not rinse well
with water, the components will be damaged.
Additional notes: Ensure that the charge contact pins
stay dry and are not touched by any metallic or greasy
parts. Do not bend the charge contact pins while they
dry. Wet charge contact pins cause an operating error
(malfunction message: the LED showing the operating
status of the charger flashes red).
3
CaviWipes™ were used for validation of the cleaning
and disinfection procedure.

Clean Charger, Handpiece, and Glare Shield

Clean all components with a soft cloth and, if necessary,
a mild cleaning agent (e. g., dish-washing detergent).
Solvents or abrasive cleaners may not be used in any
case, as they can damage the plastic components!
- Cleaning agents must not enter the unit!
To disinfect all components, spray the disinfectant on a
towel and use it to disinfect the unit. Do not spray the
disinfectant directly on the device.
- Disinfectant agents must not enter the unit!
- Dry residual disinfectants on the device with a soft
and fluff-free cloth, as they damage the plastic
components.
If necessary, ask the manufacturer of the disinfectant if
its constant use will damage plastic surfaces.
Clean the protection glass with a soft and fluff-free
cloth. Beware of scratches!
- Make sure that charge contact pins remain dry
and are not contacted by metallic or greasy parts.
Do not bend the charge contact pins during drying.
Wet charge contact pins will cause an operating
error (malfunction message: the LED showing the
operating status in the charger glows red.
11