Remove the light guide from the washer-disinfector
unit at the end of the drying cycle and store it under
low germ count conditions.
Examine the light guide for visible impurities. If any
impurities are visible, repeat the cycle in the washer-
disinfector.
Automated washer-disinfectors that meet the fol-
lowing requirements are suitable for use:
• The device generally has a certified effectiveness in
accordance with ISO 15883. The washer-disinfector
is regularly serviced and tested.
• If available, use a tested cycle for thermal disinfection
(A0 value > 600 or, for older devices, at least 5 min
at 90 °C/194 °F).
• The water used for rinsing is minimum drinking water
quality. Use low-germ and low-endotoxin water for
final rinsing.
• The air used for drying should be filtered (oil-free, with
low microbiological and particulate contamination).
Note: The evidence of suitability for automatic cleaning
and disinfection took place using the Miele G7336 CD
washer-disinfector (Gütersloh) and the cleaning agent
Thermosept RKN-zym (Schülke & Mayr). The program
D-V-THERMO0905, without drying time was used under
worst case conditions.
Cleaning
steps
Temperature:
Pre-rinse
Time:
Cleaning temperature: 45 ° ± 2 °C
Time:
Cleaning
Cleaning solution:
Concentration:
Temperature:
Final rinse
Time:
Temperature
Thermal
disinfection
Time:
Cleaning and disinfection of the light guide by
wiping is only permitted in combination with a
corresponding protective sleeve
Manual Cleaning
Clean the light guide after every application for 30
seconds with ready-to-use disinfectant wipes (e.g.,
10
Parameters
10 ° ± 2 °C
1 min
5 min
Thermosept
RKN-zym
0.3% (3 ml/l)
10 ° ± 2 °C
2 min
90 ° ± 2 °C
5 min
CaviWipes™ based on the active substances alco-
hols and quaternary compounds (QAV)) or equivalent
cleaning agents.
Follow the cleaning agent manufacturer's directions
for use.
Clean with the wipes until no contamination is visible.
Then rinse with water (drinking water quality as a
minimum) and dry with a soft, lint-free cloth.
Manual Disinfection (Intermediate Level Disinfection)
An intermediate level disinfection is possible only if
the light guide is covered with a protective sleeve
during use.
Disinfect the light guide with ready-to-use disinfec-
tant wipes (e.g., CaviWipes™ based on the active
substances alcohols and quaternary compounds
(QAV)) or equivalent disinfection agents. Follow the
disinfection agent manufacturer's directions for use,
observing in particular the contact times.
Then rinse with water (drinking water quality as a
minimum) and dry with a soft, lint-free cloth.
Manual Cleaning and Disinfection the Light Guide
by Immersion Bath
Manual Cleaning
A neutral enzymatic cleaning agent is recommended
(e.g., Cidezyme/Enzol from Johnson & Johnson).
Place the light guide for the specified application time
into the solution, making sure that it is completely
covered.
Cleaning takes place in the immersion bath with a
soft brush for at least 5 minutes.
Then remove the light guide from the solution, thor-
oughly rinse with low-germ-count water (drinking
water quality as a minimum), and dry with a lint-free
cloth.
Manual Disinfection (High Level Disinfection)
A disinfecting agent containing o-Phthalaldehyde
(e.g., Cidex OPA by Johnson & Johnson) is recom-
mended. It is compatible with the light-curing device.
To disinfect, place the cleaned light guide for the
specified application time (≥ 12 min) into the solu-
tion, making sure that it is completely covered.
The instruments in the disinfection bath must not
touch each other during this process.
Then remove the light guide from the solution, thor-
oughly rinse with low-germ-count water (drinking
water quality as a minimum) for at least 3 minutes,
and dry with a lint-free cloth.
Then check the light guide (see "Check").
1
Cleaning agent Thermosept RKN-zym (Schülke &
Mayr) for the washer-disinfector and Cidezyme
®
/

Quicklinks ausblenden: