Technische Beschreibung
Fehlersuche
Error code
Symptom
A-0B:
NE contact: Resistance too low
A-0C:
NE contact: Resistance too high
A-10:
End activation (AutoStop reached)
A-11:
No activation signal
A-12:
Activation during power up not permitted
A-17:
Simultaneous activation not permitted
A-19:
Line voltage fault present
A-1C:
Max. activation time exceeded
A-1E:
Touch screen pressed during power up
A-1F:
NE symmetry critical (cancel activation)
A-22:
AutoStop reached
A-85:
Dual-pedal footswitch disconnected
A-89:
Single-pedal footswitch disconnected
A-94:
Triple-pedal footswitch disconnected
A-95:
Instrument an MF-Buchse erkannt
A-99:
Activation type not available
A-9F:
Triple-pedal instrument disconnected, socket n
A-A3:
Footswitch not assigned to a socket
A-A4:
2 footswitches of same type connected
A-A9:
First select procedure before activation
A-AA
No mode assigned
A-AB
Instrument detection defective
A-AC
Autostart on tissue contact not possible
A-B0
NE symmetry not optimal
A-B7
Autostart only permitted up to 50W output
A-BB
SC due
A-C0
Activation type for MF-instrument missing
A-FE
Line voltage fault present
C-75
Error with instrument detection
C-84
Short-circuit between AE and NE
I-84
Defective instrument detection
I-8B
Impermissible instrument detected
I-8D
Impermissible instrument detected
I-8E
Impermissible instrument detected
all others
manuals search engine
Technical description
Troubleshooting
AUTOCON
®
II 400 Error Messages
Troubleshooting
Check neutral electrode and reposition if necessary
Check neutral electrode and reposition if necessary
End activation of HF energy
Check footswitch, handle and cable
Activate either via footswitch or handle
Check power cord, power outlet and power supply
Cancel activation
Touch screen must not be pressed during power-up
Check neutral electrode and reposition if necessary
Break contact with tissue
Reconnect dual-pedal footswitch and check cable
Reconnect single-pedal footswitch and check cable
Reconnect triple-pedal footswitch and check cable
New MF instrument at socket
Activation type not available with instrument connected
MF instrument disconnected
Assign footswitch to a socket (p. 56)
Either connect one footswitch or two of a different type
Select the required procedure (p. 52)
Assign a mode to the procedure (p. 30)
Check instrument and cable
Release tissue contact,
Instrumentbranchen ggf. trennen, Autostart zuordnen
Check neutral electrode and reposition if necessary
The max. output with Autostart should only be 50W
Arrange date for safety check
Specify activation type
Check power cord, power outlet and power supply
Check instrument and instrument cable and reconnect
No impedance modification of tissue during coagulation
Check instrument and instrument cable and reconnect
Check connected instrument and replace if necessary
Check connected instrument and replace if necessary
Check connected instrument and replace if necessary
Inform Service
Техническое описание
Поиск неисправностей
Priority
Medium
Medium
Low
High
Medium
Medium
Low
Low
Low
High
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Medium
Medium
Low
Low
Medium
Low
Low
Low
Low
Low
Low
Low
Low
Low