Herunterladen Inhalt Inhalt Diese Seite drucken

Storz AUTOCON II 400 Gebrauchsanweisung Seite 127

Inhaltsverzeichnis

Werbung

118
Downloaded from
www.Manualslib.com
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
®
AUTOCON
ll 400 model 20 5352 20-11x is intended for use in the electromagnetic environment specified below. The customer or user of the
®
AUTOCON
ll 400 model 20 5352 20-11x should ensure that it is used in such an environment.
Immunity test
Electrostatic Discharge (ESD)
IEC 61000 -4-2
Electrical fast
transient/burst
IEC 61000 -4-4
Surge
IEC 61000 -4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000 -4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000 -4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
manuals search engine
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 202
Guidance and manufacturer's declaration – electromagnetic immunity
EN/IEC 60601 test level
Compliance level
± 6 kV contact
Complies
± 8 kV air
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
Complies
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 1 kV differential mode
Complies
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Complies
<5% U
*
<5% U
*
T
T
(>95% dip in U
)
(>95% dip in U
T
for 0.5 cycle
for 0.5 cycle
Complies
40% U
40% U
T
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
Complies
70% U
70% U
T
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
Complies
<5% U
<5% U
T
T
(>95% dip in U
)
(>95% dip in U
T
for 5 sec
for 5 sec
Complies
3 A/m
3 A/m
Приложение
Указания по электромагнитной
совместимости (ЭМС)
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment. If image
distortion occurs it may be necessary to further filter
the power quality by using a AC line filter.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
T
If the user of the equipment or system requires
continued operation during power mains interruptions,
it is recommended that the equipment or system be
powered from an uninterruptible power supply or a
)
battery.
T
)
T
)
T
Power freq uency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis