Validated reprocessing procedures
Overview of the reprocessing procedures of the
components
Category
Surface disin-
Machine clean-
fection with
ing with thermal
cleaning
disinfection
Flow tube
Yes
No
AGS head-
No
Yes
er
Container
No
Yes
for buffer
volume
Transfer
No
Yes
hoses and
scaveng-
ing hoses
Plugs and
No
Yes
adapters
Surface disinfection with cleaning
Categories:
Flow tube
Surface disin-
Manufac-
Con-
fectant
turer
centra-
tion
Dismozon plus
BODE
1.6 %
Chemie
Neodisher LM2
Dr. Weigert 2 %
Prerequisites:
The surface disinfectant has been prepared in
accordance with the manufacturer's instructions.
The manufacturer's instructions, e.g., regarding
shelf life or application conditions, are observed.
An uncontaminated, lint-free cloth soaked in sur-
face disinfectant is used for the cleaning surface
disinfection.
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Device malfunctions
Ensure that no liquid penetrates the device.
Cleaning
1
Wipe off obvious soiling with a disposable cloth
soaked in surface disinfectant. Dispose of the
cloth.
2
Wipe all surfaces. After that, there must no lon-
ger be any soiling visible.
Surface disinfection
3
Wipe cleaned surfaces again to visibly wet all
surfaces to be disinfected with surface disinfec-
tant.
4
Wait for the surface disinfectant contact time.
5
At the end of the contact time, moisten a new,
uncontaminated, lint-free cloth with water (at
least drinking water quality).
6
Wipe all surfaces until no remains of the surface
disinfectant, such as foam residues or streaks,
are visible.
7
Wait until the surfaces are dry.
8
Check the surfaces for visible damage and, if
necessary, replace the product.
Supplementary information
Manual precleaning (AGS header only)
1
Prepare the cleaning solution in accordance with
the manufacturer's instructions.
2
Immerse the product in the solution, making
sure to displace trapped air.
3
At the beginning and towards the end of the con-
tact time, swirl the product backwards and for-
wards several times in the solution. Make sure
that the solution reaches all surfaces and interior
spaces.
Instructions for use Anesthetic gas receiving system AGS
4
After the contact time has elapsed, rinse the
product under running water until residual clean-
ing agent is no longer discernible.
5
Check the product for visible damage and re-
place if necessary.
6
Thoroughly shake out residual water.
 Refer to the sections "Preparation" and "Opera-
tion", see page 9.
Machine cleaning with thermal
disinfection
Use a washer-disinfector that meets the require-
ments of the standard ISO 15883. Dräger recom-
mends the use of a load carrier for anesthesia ac-
cessories and ventilation accessories. Follow the
manufacturer's instructions for the washer-disinfec-
tor.
Components:
AGS header
Container for buffer volume
Transfer hoses and scavenging hoses
Plugs and adapters
Contact
Step
Agent
time
15 min
Prelimi-
Tap wa-
nary
ter
10 min
cleaning
Cleaning
Neo-
disher
Medi-
clean
forte
Neutrali-
Neo-
zing
disher
Z
Flushing
Demi-
nera-
lized
water
Disinfec-
ting
Drying
Prerequisites:
The washer-disinfector has been prepared in accor-
dance with the manufacturer's instructions.
Positioning the components in the load carrier
Procedure:
1
Position the components to be stable.
2
Ensure the following:
All surfaces and interiors can be completely
rinsed.
The water can drain off freely.
Performing reprocessing
1
Select a cycle.
2
When the cycle has ended, check the compo-
nents for visible soiling and repeat the cycle if
necessary.
3
Check the components for visible damage and
replace if necessary.

Storage and transport

After reprocessing, there are no special require-
ments for storage and transport of the product. How-
ever, the following must be observed:
Store dry and free of dust
Avoid recontamination and damage during
transport
Man-
Con-
Tem-
Con-
ufac-
cen-
pera-
tact
turer
tra-
ture
time
tion
Tap wa-
Min.
ter tem-
2 min
perature
Dr.
Min.
Wei-
5 min
gert
Dr.
Min.
Tap wa-
Min.
Wei-
0.1
ter tem-
1 min
gert
%
perature
Tap wa-
Min.
ter tem-
1 min
perature
Min.
Min.
93 °C
5 min
(199.4 °
F)
Drying
time
de-
pends
on the
load
English/English US
All further information on storage and transport in-
cluded in the accompanying documents must be ob-
served.
Other agents
Disinfectants
Use nationally approved disinfectants suitable for
the respective reprocessing process and the intend-
ed application.
Surface disinfectants
The manufacturers of the surface disinfectants have
verified at least the following spectra of activity:
Bactericidal
Yeasticidal
Virucidal or virucidal against enveloped viruses
Follow the manufacturer's instructions for surface
disinfectants.
The following surface disinfectants were compatible
with the material at the time of testing:
Class of
Surface dis-
Manufac-
active in-
infectant
turer
gredient
Chlorine-
BruTab 6S
Brulin
releasing
Clorox Pro-
Clorox
agents
fessional Dis-
infecting
Bleach
Cleaner
Dispatch
Hospital
Cleaner Dis-
infectant Tow-
els with
Bleach
Klorsept 17
Meden-
tech
Actichlor plus Ecolab
USA
Oxygen-re-
Descogen
Antisepti-
leasing
Liquid
ca
agents
Descogen
Liquid r.f.u.
Oxygenon
Liquid r.f.u.
Dismozon
BODE
plus
Chemie
Oxycide
Ecolab
USA
Perform
Schülke &
Mayr
SteriMax
Aseptix
Wipes
Incidin Oxy-
Ecolab
Wipes
USA
2)
Quaterna-
acryl-des
Schülke &
ry ammoni-
Mayr
Mikrozid alco-
um com-
hol free liq-
pounds
2)
uid
Mikrozid alco-
hol free
2)
wipes
Mikrozid sen-
2)
sitive liquid
Mikrozid sen-
2)
sitive wipes
Cleanisept
Dr.
Wipes Maxi
Schum-
acher
Surfa'Safe
ANIOS
Premium
Laborato-
ries
Wip'Anios Ex-
cel
Tuffie 5
Vernacare ARTG
1) United States Environmental Protection Agency
2) Virucidal against enveloped viruses
3) Australian Register of Therapeutic Goods
Dräger states that oxygen-releasing agents and
chlorine-releasing agents may cause color change
in some materials. Color change does not indicate
that the product is not functioning correctly.