5.4 Static Alignment
+ = anterior placement / – = posterior placement (in relation to the load line)
No.
see fig. 6
Required equipment:
743L100 L.A.S.A.R. Posture
   
To determine the load line
Prosthetic foot (with shoe) on force measurement plate (sufficient load: > 35% body weight)
Other foot (with shoe) on height compensation plate
The tips of the shoes are in one line
   
Optimise the static alignment solely by changing the plantar flexion. Adjustment is performed only
with the anterior and posterior set screws of the tube adapter on the prosthetic foot.
a–p positioning of the alignment reference point
the load line:
-35 mm
5.5 Dynamic Trial Fitting
CAUTION
Adjusting the Settings
Fall due to incorrect or unfamiliar settings
Only adapt the settings to the patient gradually.
Explain the effects of the adjustments on the use of the prosthesis to the patient.
During dynamic trial fitting, the alignment and settings of the prosthesis are checked and adapted
for optimum walking according to the needs and abilities of the patient.
The patient has to learn the safe use of the prosthesis through intensive training.
The following subsections describe the adjustment possibilities for adapting the product to the
patient.
The list below provides an overview of the sequence for working through the subsections:
Checking the Factory Settings
Setting Flexion Damping
Setting Extension Damping
Setting Stance Phase Damping
Setting the Stance Phase Damping Threshold
The adjustment is illustrated graphically in the supplied Quick Start Guide.
5.5.1 Opening the Covers
Tools required:
>
Flat screwdriver
1) Open the extension damping and flexion damping covers (see fig. 11).
2) Open the stance phase damping cover (see fig. 12).
5.5.2 Checking the Factory Settings
Function
Flexion damping
Extension damping
Stance phase damping
Static Alignment Process
, position the patient on the L.A.S.A.R. Posture as follows:
Means of adjustment
Adjustment valve "F"
(see fig. 13)
Adjustment valve "E"
(see fig. 14)
Adjustment ring "S"
(see fig. 15)
(rotation axis of the prosthetic knee joint) to
Factory setting
Stop "–"
Stop "–"
Stop "+"
Meaning
Minimum damping
Minimum damping
Maximum damping
Ottobock | 27