Select an X-ray intensity level.
6.
Category
Child
(Age 12 or
under)
Adult
(Above Age
Once the X-ray intensity level is selected, the system automatically sets the
7.
exposure condition. To change the condition. Click the plus-minus button next
to the voltage and the current symbols.
Each time you click the button, the voltage value changes by ± 1.0 kVp, and
the current value by ± 0.1 mA.
The table below shows the range of exposure conditions that you can adjust.
Mode
CBCT
Click Confirm. After clicking the button:
8.
▪
The rotating unit moves to the scanning position.
▪
Laser beams are turned on.
▪
The progress bar appears above the Confirm button.
▪
The scan information window shows the estimated DAP (Dose Area
Product), scan time, and exposure time.
Green X 12
TM
(Model: PHT-75CHS) User Manual
Average head
circumference (cm)
53±3
56±3
12)
Minimum Value
kVp
60
6. How to Acquire CBCT Images
Range (cm)
>53±3
53±3
<53±3
>56±3
56±3
<56±3
Maximum Value
mA
kVp
4
99
Intensity level
Hard
Normal
Soft
Hard
Normal
Soft
kVp
14
87