How To Acquire Cbct Images; Preparing X-Ray System - Vatech Green X 12 Bedienungsanleitung

Inhaltsverzeichnis

6. How to Acquire CBCT Images

How to Acquire CBCT Images
6.

Preparing X-ray System

6.1
Follow the steps below to prepare an image capture.
Clean all accessories that come into direct contact with a patient, with a
1.
chlorine dioxide-based and alcohol-free product.
Insert a bite block into the chinrest. Ensure to select the bite block suitable for
2.
your purpose.
Insert left and right temple supports.
3.
Put a sanitary vinyl cover on the bite block. Always change the cover after
4.
each scanning.
76
For disinfection, the accessories that come into
direct contact with a patient must be cleaned
before scanning.
Bite Block
Normal Bite
Special Bite B
Special Bite A
Green X 12
Purpose
For the normal mode and non-
edentulous patients.
For the normal mode and edentulous
patients.
For the special mode (TMJ and Sinus).
This bite is used for both edentulous
and non-edentulous patients.
Sanitary Vinyl Cover
Normal Bite
(Model: PHT-75CHS) User Manual