4.4 CALIBRATION MENU

ENTER
Press
The internal Switch 2 must be set in the unlocked position to enter parmeters within this
menu. (section 3.4.1)
The following parameters can be set within this menu:
LANGUAGE
K-FACTOR
CURRENT
0......9
FREQUENC
The following sections explain how to change the parameter values within the calibration
menu above.
8045
CALIBRATION
simultaneously for 5 seconds.
Function
Selection between English, German,French,
Italian and Spanish.
UNIT
Selection of engineering units to displayflowra-
te and totalizer.
Calculation of the process K-factor.
Determination of 4-20 mA measuring range.
PULSE
Parameter definition of pulse output (unit and
volume).
RELAY
Parameter definition of relays. This message
only appears on devices with relay outputs.
FILTER
Damping selection. There are 10 steps availab-
le and 2 filtering modes.
TOTAL
Totalizer resetting.
Selection of the mains AC frequency
(50 or 60 Hz).
END
Return to operation mode and storage of new
parameters.