INSTALLING THE SUPPORT SURFACE
WARNING
• The risk of entrapment can develop when the support surface is placed on bed frames that leave gaps of even a
few inches between the support surface and the headboard, footboard, and siderails . The support surface is not
to be used when such gaps are present .
• To avoid the risk of patient injury, do not use the unit on a bed frame of a larger or smaller width than the stated
width as this may cause the support surface to slide .
Prerequisite: Per OSHA, two (2) operators are required to setup the support surface .
1 .
Ensure that the support surface properly fits the bed frame on which it is being placed .
2 .
Ensure the dedicated sloped heel section is installed at the foot end of the bed frame .
3 .
Place the linens on the support surface as loose as possible to avoid creating a "hammocking" effect and
increasing interface pressures .
CAUTION
To avoid the risk of equipment damage, do not put overlays or accessories inside the cover . Doing so may reduce
pressure redistribution performance .
www.stryker.com
Setup Procedures
2860-009-005 REV B
English
Return To Table of Contents
1-11