Taking a Bilateral Pupil Measurement
Two components are required to initiate a bilateral pupil measurement:
NPi-300 Pupillometer (Ex. 12)
Single Patient Use SmartGuard (Ex. 13)
Open a new SmartGuard. Slide the SmartGuard onto the NPi-300 with the
foam pad at the bottom (Ex. 12). There will be an audible click when the
SmartGuard is properly positioned.

1. Associate SmartGuard with Patient ID

For the first patient use, the SmartGuard requires a one-time association
of the Patient ID. For subsequent measurements, the Patient ID will be
permanently saved on the SmartGuard, which can store up to 168 bilateral pupil
measurements of the associated patient.
There are two options for associating the Patient
ID to the SmartGuard. Select Scan Code to scan
the patient's wristband barcode using the NPi-
300 Incorporated Barcode Scanner, or select
Manual ID to manually enter the Patient ID with
either alpha or numeric characters (Ex. 14).
Scan Barcode Using Incorporated Barcode
Scanner
Select Scan Code. The NPi-300 will emit a white
light from the top of the device (Ex. 15). Center
the light over the barcode until you hear an
audible beep (Ex. 16). The Patient ID will now
appear on the NPi-300 touchscreen. Confirm the
patient information is correct and select Accept
(Ex. 17). The NPi-300 will display the Patient ID
and read Ready to Scan (Ex. 18).
Manual Entry of the Patient ID
Select Manual ID. Using the touchscreen or keypad, enter the alpha or numeric
Patient ID and select
is correct and select Accept (Ex. 17). The NPi-300 will display the Patient ID and
read Ready to Scan (Ex. 18).

2. Measure Pupils

Position the NPi-300 with SmartGuard at
a right angle to the patient's axis of vision,
minimizing any tilting of the device (Ex. 20).
NeurOptics® NPi®-300 Pupillometer System—Instructions for Use ©2021 NeurOptics, Inc.
(Ex. 19). Confirm the patient information on the screen
Ex. 12
Ex. 15
Ex. 17
Ex. 13
Ex. 14
Ex. 16
Ex. 18
Ex. 19
Ex. 20
6