Handling, Cleaning and Maintenance

Always handle the NPi-300 Pupillometer and NPi-300 Charging Station with care because sensitive metal,
glass, plastic and electronic components are contained inside. The NPi-300 and Charging Station can be
damaged if dropped or by prolonged exposure to liquid or high humidity environments.
The NPi-300 and Charging Station do not require any regularly scheduled maintenance. If the NPi-300 and
Charging Station are not working properly, or are believed to have been damaged, immediately contact
NeurOptics Customer Service at
Info@NeurOptics.com.
250-9792, or email:
Cleaning the NPi-300 Pupillometer and NPi-300 Charging Station
Isopropyl alcohol (IPA)-based cleaning solutions, in formula concentrations up to 70% IPA, are recommended
for use in cleaning the NPi-300 and Charging Station. Do not use chemicals that can damage the NPi-300 and
Charging Station surface. Some chemicals can weaken or damage plastic parts and may cause instruments to
not operate as intended. Use all cleaning products per manufacturer's instructions, being careful to squeeze
out excess liquid prior to wiping the NPi-300 and Charging Station and do not use an oversaturated cloth.
Wipe all exposed surfaces. Follow the cleaner's manufacturer instructions as to the time required to leave the
solution on the device surface.
• DO NOT
use an oversaturated cloth. Be sure to squeeze out excess liquid prior to wiping the NPi-300 or
the Charging Station.
• DO NOT
allow the cleaner to collect on the instrument.
• DO NOT
use any hard, abrasive or pointed objects to clean any part of the NPi-300 or Charging Station.
• DO NOT
immerse the NPi-300 or the Charging Station in liquid, or attempt to sterilize the product, as
damage to the electronic and optical componentry could occur.
Drying and Inspection Following Cleaning
Confirm the NPi-300 and Charging Station are thoroughly dry before placing the NPi-300 back into the
Charging Station.
Cleaning Considerations: NPi-300 Liquid Crystal Display (LCD)
For best protection of the liquid crystal display (LCD), use a clean, soft, lint-free cloth and up to 70% IPA to
clean the NPi-300 optics. Occasional cleaning of the Incorporated Barcode Scanning Window (located just
above the NPi-300 Lens) is also recommended using a clean, soft, lint-free cloth and up to 70% IPA.
In instances where there is concern of exposure to highly resistant bacteria, viruses, fungi or spores (ie:
Clostridium difficile, or "C. diff"), we understand that hospital protocols may require use of cleaning
solutions containing sodium hypochlorite (bleach) when cleaning equipment. If products containing sodium
hypochlorite (bleach) are used to clean the LCD of the NPi-300, the cleaning process should be followed by
a second cleaning using a clean, soft, lint-free cloth and up to 70% IPA to ensure that all bleach residue is
completely removed from the LCD.
NeurOptics® NPi®-300 Pupillometer System—Instructions for Use ©2021 NeurOptics, Inc.