More Settings
Customize Barcode Scanner
The NPi-300 Incorporated Barcode Scanner can be customized to truncate
or expand the alpha or numeric characters read from a hospital barcode, if
required. The Default settings automatically adjust to read most types of
1D and 2D hospital barcodes, and "Default" should remain selected unless a
specific customization needs to be applied to all barcodes scanned by the
NPi-300. Select Custom Barcode
scan a sample barcode and program the required customizations (truncation
or expansion) to be used for all future scans. Contact NeurOptics for
additional information.
System Information
Select System
(Ex. 43) to view the NPi-300's system information, displaying the Serial Number, Software
Application, and Firmware versions of the device.

Troubleshooting

Issue
1. NPi-300 Pupillometer will
not turn on
2. Pupil measurement will not
initiate after release of the
RIGHT or LEFT key
3. NPi-300 returned to Home
Screen while taking a
measurement
4. "Rescan" displayed
following measurement

Power Off

To turn the NPi-300 Pupillometer OFF, either:
Navigate to the Home Screen, and select the Power icon
confirm Yes to power OFF (Ex. 44).
Press and hold the On/Off button
The NPi-300 may occasionally require a System Reboot. To reboot, simply
press and hold the On/Off button
powers OFF.
NeurOptics® NPi®-300 Pupillometer System—Instructions for Use ©2021 NeurOptics, Inc.
(Ex. 43), then select Scan Sample to
Possible Reason
Using incorrect Power Adapter
Power cord is not fully plugged into
the wall or the charging station
Battery completely discharged
Too much blinking
Device not held correctly
RIGHT or LEFT button was
pressed while measurement
was being completed, causing
measurement to be aborted
NPi-300 is moved from position
prior to completion of the
measurement
Patient blinked during
measurement
on the side of the NPi-300.
on the side of the NPi-300 until it
Solution
Use only Power Adapter provided with
NPi-300. Check label on power adapter.
Check connections.
Charge the battery by placing the NPi-300
into the Charging Station.
Gently hold patient's eye open with your
finger during measurement.
Hold device at a 90-degree angle to
patient's face. Make sure patient's pupil is
centered on the screen.
Repeat the scan, ensuring no buttons are
pressed until the scan is completed and
results appear on the screen.
Repeat the scan and maintain proper
position of the NPi-300 until the
measurement is completed and pupillary
measurements are displayed.
Hold the patient's eyelid open and repeat
the scan.
, then
Ex. 43
Ex. 44
11