such behavior occurs, the user should try to move the interfering equipment
further from this device.
Make sure the A-71100-XX/ A-71101-XX operate correctly when used near
other electronic devices. Portable and mobile radio frequency (RF)
communications equipment can affect electrical equipment.
Medical equipment needs special precautions in regard to electromagnetic
compatibility (EMC) and needs to be installed and put into service
according to the EMC information supplied in the tables that follow.
Guidance and Manufacturer's Declaration—Electromagnetic
The A-71100-XX/ A-71101-XX are intended for use in the electromagnetic environment
specified below.
The customer or the user of the A-71100-XX/A-71101-XX should make sure it is used in
such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The A-71100-XX/ A-71101-XX are intended for use in the electromagnetic environment
specified below. The customer or the user of the A-71100-XX/A-71101-XX should make
sure it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Allen® Advance Table User Manual (80028190)
Emissions
Compliance
Electromagnetic Environment—Guidance
Group 1
The A-71100-XX/A-71101-XX use RF energy
only for its internal functions. Therefore, its RF
emissions are low and are not likely to cause any
interference in nearby electronic equipment.
Class A
The A-71100-XX/A-71101-XX are suitable for
use in all establishments, other than domestic,
and those directly connected to the public low-
voltage power supply network that supplies
Class A
buildings used for domestic purposes.
Complies
IEC 60601
Test Level
± 6 kV Contact
± 8 kV Air
± 2 kV (power line)
± 1 kV
(Input/Output line)
Compliance
Level
± 6 kV Contact
± 8 kV Air
± 2 kV (power line)
± 1 kV
(Input/Output line)
55
SPECIFICATIONS
Electromagnetic
Environment—
Guidance
Floors should be wood,
concrete, or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.