Persons a nd equipment will not be in the way of the support tops
when they get flipped.
To watch the lines so that they don't get pinched or caught during
the flip.
The patient's head stays in the correct position during the flip.
–
Do as follows to unlock the
support tops for the flip:
a. Lift the end of the flip handle (A)
until it locks in position.
10. Push and hold the red safety lever (B)
forward as you pull the flip handle (A)
down to the released position.
NOTE:
The manual flip indicator on the
display turns yellow to show that the
support tops are no longer locked in
position and they can be flipped. The
pendant indicator on the display turns
off to show that you can not use the
pendant controls while the Allen
Advance Table is in flip mode.
Allen® Advance Table User Manual (80028190)
Instructions for Use