Functionality Tests
Installation report
Extensive
functionality test
Test Intervals
Execution
KNAUER
42
Functionality Tests
Installation Qualification (IQ)
Certification on the functionality of the device. During installa-
tion of the device, an installation report (IQ document) is cre-
ated upon request in coordination with the Technical Support of
KNAUER.
This installation report needs to be completed in full and signed
by both parties. It serves as proof of the properly executed
installation and functionality of the device.
Operation Qualification (OQ)
Extensive test of the device's functionality. A successfully exe-
cuted OQ ensures that the device functions properly.
Run the extensive functionality test at the following time inter-
vals:
Average use
1 to 5 days/week:
More than 5 days/week or 24 hours/day:
Operation with buffer solutions or other salt
solutions:
The test can carried out either by KNAUER's Technical Support
or by a technical service authorized by KNAUER.
The OQ documentation required for executing the OQ is with
costs (once) and can be ordered separately from KNAUER.
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Device test
Every 6
months
Every 3
months
Every 3
months