Herunterladen Inhalt Inhalt Diese Seite drucken

Integrating The Detector Into A System; Pre-Installed Capillaries; Fig. 21 Installation Of The Detector - Knauer Azura DAD 2.1L Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für Azura DAD 2.1L:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

30
Integrating the Detector into a Sys-
tem
To integrate the pump into a system, note the ambient condi-
tions found in the sections Operating Environment and Techni-
cal Data as well as the ambient conditions of other devices to be
integrated into that system.
The detector is integrated into the HPLC flow system by con-
necting the capillary to the flow cell and the HPCL system.
Procedure
Stainless steel fittings are tightened with 5 Nm, PEEK fittings are
tightened with 0.5 Nm.
Capillary layout

Pre-Installed Capillaries

Pre-installed stainless steel and PEEK capillaries are color-coded
according to their inner diameter.
Note
PEEK capillaries are not suitable for use with pure Acetonitrile.
Acetonitrile can cause capillaries to crack or rupture.
KNAUER
Installation
Fig. 21
Installation of the detector
Color
Material
red marker
Stainless steel
blue marker
Stainless steel
black marker
Stainless steel
red stripes
PEEK
blue stripes
PEEK
orange stripes
PEEK
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Inner diameter
0.1 mm
0.25 mm
0.4 mm
0.1 mm
0.25 mm
0.5 mm

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Azura mwd 2.1l

Inhaltsverzeichnis