Installation

Integrator Connector

The integrator connector sends measuring signals from the
detector.
non-bipolar
4 channel
0 to 5 V
DAC 16 bit
scalable
Adjustable to offset

Power Supply

Note
The nominal capacity of the connected devices must be maxi-
mum 50 % of the power supply to account for larger inrush cur-
rents when switching on the modules.
The maximum power input is 75 watts.
Prerequisites
The electrical power supply at the installation site must be
connected directly to the nearest main power line.
The power must be free from ripple, residual current, voltage
peaks and electromagnetic interference.
The connectors for the mains voltage are grounded accord-
ingly.
The device receives sufficient power with reserve capacity.
The device is intended for use with AC power networks of 100 –
240 V.
Power cable
Use only the enclosed power cable to connect the device to the
mains. Replace defective power cables only with accessories
from KNAUER. Only use power cables admitted for use in your
country.
In case of queries, contact the Technical Support.
Power plug
Make sure that the power plug on the rear of the device is
always accessible, so that the device can be disconnected from
the power supply.
Next steps
Finish the installation and connect the device to the mains.
KNAUER
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
CAUTION! 
Electric hazard if the power adapter is turned on
while connecting or interrupting the power con-
nection.
Switch off the power adapter beforehand.