48
Troubleshooting
System message
"Spectrum buffer
overflow"
"Spectrum output
busy or not ready"
T
"Temperature con-
trol failed"
"Time already
exists"
W
"Wake up time
already passed"
"WL/BW out of
spectral range"
"WL validation
failed"
KNAUER
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Error List and Solutions
Internal 3D data buffer is exhausted
due to LAN connection failure.
3D data acquisition cannot be started.
Wait until the data transfer is com-
plete before continuing.
Maximum temperature on lamp unit
exceeded.
Correct the time entry.
The wake up entry is set in the past.
Enter a new time.
Selected wavelength and bandwidth
exceed the spectral range. Extend the
selected spectral (scan) range or
change the selected wavelength and /
or bandwidth.
The wavelength precision validation
has failed. Restart the validation. Hol-
mium oxide and/or H
specification.
, H
lines out of