50
Contact the Technical Support
If you have any technical questions regarding KNAUER hardware
or software, please use one of the contact options below:
Technical Support hotline:
Languages: Available by telephone
European hotline
in German or English:8 am to 5 pm (CET)
Phone:+49 30 809727-111
Telefax:+49 30 8015010
E-mail contact:
support@knauer.net
Maintenance Contract
The following maintenance work on the device may only be per-
formed by KNAUER or a company authorized by KNAUER and is
covered by a separate maintenance contract:
• Opening the device or removing housing parts.
Maintenance Intervals
Operating hours
The Control Unit, the Mobile Control and the different software
products (e.g. ClarityChrom
play or read out the operating time of the detector. You find a
detailed description on how to display or read out GLP data in
the respective user manuals.
Cleaning and Caring for the Device
All smooth surfaces of the device can be cleaned with a mild,
commercially available cleaning solution, or with isopropanol.
KNAUER
Maintenance and Care
Operating hours
1000
2000
CAUTION!
Intruding liquids can cause damage to the device!
Place solvent bottles next to the device or in a
solvent tray.
Moisten the cleaning cloth only slightly.
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
®
®
or OpenLAB
) enable you to dis-
Measures
Replace halogen lamp (DAD 6.1L
only)
Replace deuterium lamp