Communication
Technical parameters
Ambient conditions
General
Detection
KNAUER
Technical Data
Drift
Linearity
Time constants
Integration time
Maximum data rate
Inputs
Outputs
Analog outputs
Control
Programming
GLP function
Display
Temperature range
Humidity
Power supply
Maximum power
consumption
Dimensions
Weight
Leak sensor
Degree of protection
Height above sea
level

MWD 2.1L

Detector type
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
-4
AU at 254 nm 
5.0 x 10
(ASTM E1657-94)
> 1.8 AU
0 / 0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5
/ 1.0 / 2.0 / 5.0 / 10.0 s
Automatic
100 Hz (LAN)
Error (IN), Start (IN), Autozero
Events 1–2, Error (OUT), + 5 V, 
Valve + 24 V, Valve (OUT)
4 × 0-5 V scalable, 16 bit, offset
adjustable
Digital: LAN-DHCP, connector
Manual: Mobile Control (optional)
Wake-up
Detailed report including lamp recog-
nition, operating hours, lamp operat-
ing hours, number of lamp ignitions.
Mobile Control
4–40 °C, 39.2–104 °F
below 90 %, non-condensing
Input 100–240 V
Output 50–60 Hz
75 Watt
361 mm x 158 mm x 523 mm
(Width × Height × Depth)
13.8 kg
Yes
IP 20
maximum 2000 meters
Variable multi-wavelength UV/VIS
detector

Werbung