Fig. 19 Relation Between Display And Events Terminal Strip - Knauer Azura DAD 2.1L Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für Azura DAD 2.1L:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

26
Installation
For sending start, control and error signals to external devices
Fig. 19
Relation between display and Events terminal strip
Explanations for Pin Header Connector
Connection
(Event 1)
EV 2 (Event 2)
Error OUT
KNAUER
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Function
Relay contact
EV 1
The contact is on a floating basis. Its
setting depends on the settings in the
Control Unit, the Mobile Control, or
software.
Steady-rate signal:
passive = open relay contact
active = closed relay contact
Pulse:
Closed relay contact for at least
1000 ms
Permissible load of the relay contact:
1 A/ 24 V DC
TTL output
Levels:
passive 5 V
active 0 V
Pulse:
0 V for at least 1000 ms
TTL output
Levels:
passive 5 V
active 0 V
Output is active until the Error con-
dition has been eliminated.