Herunterladen Inhalt Inhalt Diese Seite drucken

Fig. 33 Lamp Cover; Fig. 34 Lock Ring Of The Lamp Plug; Fig. 35 Screws On The Lamp Socket - Knauer Azura DAD 2.1L Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für Azura DAD 2.1L:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

54
error message is displayed. In addition, the red LED lights up
and the center LED does not light.
Procedure
Deuterium lamp
Next steps
Insert a new lamp.
KNAUER
Maintenance and Care
DANGER! Risk of death due to high voltages.
Do not change lamps if the lamp area is not volt-
age-free.
Check the status of the lamps in the software and
on the LEDs.
WARNING! Burn hazard from hot lamp! 
Switch off the lamp meanwhile the device
remains switched on for further 15 minutes, oth-
erwise the lamp can not cool down. 
Afterwards, switch off the device and pull the
power plug.
Process
1. Loosen the screws
1
and
of the lamp
2
cover.
2. Remove the lamp
cover.
3. Loosen the lock ring
3 or 4 of the lamp
plug and pull out
the plug.
4. Using the screw-
driver, loosen the
two screws
or
5
6
5. Remove the lamp.
DAD 6.1L, DAD 2.1L, MWD 2.1L User Manual V6700, Version 1.0
Figure
Fig. 33
Lamp cover
3
4
Fig. 34
Lock ring of the lamp plug
.
Fig. 35
Screws on the lamp socket
1
2
5
6

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Azura mwd 2.1l

Inhaltsverzeichnis