1.3
Features / limits
The following features depend on the Field Bus Cards used, the addressing / configuration and the available
software licenses:
•
Modular replacement of field buses
•
SD card slot for future firmware updates
•
Integrated firewall
•
Option for checking the power supply (local and station)
•
Communication (data / audio) with other controllers via ETH-LAN or the zone data bus (POF)
•
Centralised and decentralised installation
•
Call answering via DECT and ISDN standards
•
Operation as master / slave controller possible
•
Centralised operation
•
Service functions such as:
Firmware and software updating for system devices
Remote maintenance and network diagnosis function (SNMP) with PC
Basic commissioning
Status LEDs
•
Backwards-compatible with support for the functions of zone controller IP as of V05.10
•
Same functional concept for zone controllers and SMCs in Clino System 99plus
•
Stand-alone operation (without field bus card POF-POF) possible as of firmware V05.11R000
Performance Limits
Number of zone databus devices
Length of zone databus
Number of corridor databus devices
Number of corridor databus plus
bed databus devices
Length of corridor databus / audio bus
Number of bed databus devices
Length of bed databus
If the SMC is used for system migration, always take note of the performance limits of the
corresponding system.
FB 89470B1 / 10.13
Ward Management Controller Clino System 99plus (SMC99plus)
:
max. 75 devices (of which max. 64 SMCs)
:
max. 1000 m (max. 50 m without fibre-optic converter)
:
max. 127 devices (repeaters required as of 64)
:
max. 255 devices
:
max. 2500 m (max. 500 m without repeaters)
max. 10 devices, e.g. 8 bed modules or passive beds,
:
1 display module, 1 RFID module
:
max. 200 m
45