TO ADJUST THE ACTIVE THERAPY PARAMETERS

1 .
Select the "Percussion" button .
2 .
Adjust the therapy parameter as described in
step 2 of section "To Start Percussion Therapy"
on
page
1-19 .
Note:
• The adjusted parameters take effect as soon as
they are changed .
• To change lung selection, Percussion must be
stopped .
• The percussion therapy session continues until
the selected duration time is reached or until the
session is stopped or paused .
• When restarting percussion therapy after the
session was paused or stopped for more than
30 seconds, the screen will display two options:
"Continue Previous Therapy Session" or "Start New
Therapy Session" .
• All therapy parameters are stored until changed or
reset .
• To reset the parameter to the default values, refer to
"Advanced Settings" on
• The vibration therapy button is inactive while
Percussion is active .

TO STOP PERCUSSION THERAPY

1 .
Select the "Stop" button and the percussion therapy
is stopped .
Return To Table of Contents
1-20
Operation Guide
page 1-29 .
2950-109-005 REV B