SERVICE SCREEN FUNCTIONS

To run the support surface diagnostic test, select the
"Diagnostic Test" button . The test runs automatically
and takes a few minutes . The support surface bladders
inflate and are tested for leaks .
WARNING
Do not perform the Diagnostic Test with a patient on the
support surface to ensure patient safety .
Note: Do not start the diagnostic test unless the cells
have been deflated . An external pressure, such as
leaning on the support surface or an external piece
of equipment in contact with the support surface can
cause the diagnostic test to report inaccurate results .
"Burn In" is used in 'Manufacturing' to test the support
surface prior to shipment . Runs functions automatically
for 24 hours .
WARNING
Do not perform Burn In function in the field, this function
is for 'manufacturing' purposes only .
Press the "Restart Log" button to update event information
from the CPU . Use when replacing a controller .
Return To Table of Contents
1-56
Service Information
2950-109-005 REV B
www.stryker.com