ALARMS / ERRORS

If the screen displays "ALARM Service Required by
Qualified Service Personnel", something has occurred
to prevent the support surface from providing therapy .
Service is required by qualified service personnel .
• If the message displays when the support surface is
powering up, remove the support surface from use
and notify service .
• If the message displays when a therapy is selected,
service will be required before the selected therapy
can be used .
If the screen displays "ALARM Auto . Overload Caused
a Power Shut Down Ref . Oper . Manual", the support
surface will automatically shutdown . Immediately unplug
the support surface power cord from the power source
and remove the patient from the support surface .
• After 30 minutes have elapsed, if the condition
has cleared and the unit is plugged into the power
source, the support surface will restart . If the
overload condition occurs again within 24 hours,
the support surface cannot be used until service is
performed by qualified service personnel .
ERRORS
Service Event
MAIN_LEAK
LEFT_SURFACE_LEAK
RIGHT_SURFACE_LEAK
LEFT_PERCUSSION_LEAK
RIGHT_PERCUSSION_LEAK
INVALID_COMMAND
www.stryker.com
Service Information
Code
136
Diagnostic test leak rotation bladder circuit .
137
Diagnostic test leak rotation bladder circuit .
138
Diagnostic test leak rotation bladder circuit .
139
Diagnostic test leak percussion bladder circuit .
140
Diagnostic test leak percussion bladder circuit .
146
An invalid command has been received by the
support surface controller .
2950-109-005 REV B
Description
Return To Table of Contents
English